Custom Sensor Solutions, Inc.

Specifications of the Model 1010 Precision Gas Diluter

FeatureSpecification
Dilution range2% to 100% (50-fold to 0-fold)
Accuracy2% of final concentration over 5% to 100% range
Flow rate0 to 600 cc/min (can be used at high flows with simple modification)
Inlet pressuresatmospheric only at both ports
Materials contacting gas

stainless steel, polypropylene, polyethylene, Viton® (in valve seal)

graphite composition (pumped version only)

Compatible gases (Note 1)

all concentrations of air, oxygen, nitrogen, carbon dioxide, noncondensing water vapor, and inert gases

carbon monoxide (0.1 to 10,000 ppm)

nitrogen dioxide (dry only, 0.005 to 100 ppm)

nitric oxide (dryonly, 1 to 100 ppm)

hydrogen sulfide (dry only, 0.005 to 100 ppm)

sulfur dioxide (dry only, 1-500 ppm)

ethylene oxide (0.1 to 10,000 ppm)

most hydrocarbons and chlorohydrocarbons above 1 ppm

Incompatible gases (Note 1)

hydrogen chloride, bromide, iodide, fluoride (all concentrations)

combustible mixtures of all gases

benzene and other VOC's below 0.1 ppm

Size2.6" H x 8.2" W x 7" D

Note 1: The list of compatible and incompatible gases is not exclusive. The list of compatible gases includes only those that have been tested by us for extended periods.

Note 2: Viton® is a registered trademark of DuPont Dow Elastomers

Current Feb. 23, 1998